3. Press in the
POWER
switch [1J. About 30 seconds later, rotate the
INTEN
[3] control clockwise until tbe trace
app
ears.
And
then,
adjust brightness so as to
mak
e it suitable for observing.
<Caution> A burn-resistant material is used in the
CRT
. However if
the CRT is left with an extremel y bright dot or trace for
a very long time, the screen may be damaged. Th erefore,
if a measurement requires high brightness, be certain to
turn down the I
NT
EN control immediately afterward. Also,
get in the
hab
it of turning the brightness way down if
the scope is
left unattended for any period of time.
4. Adjust the
FOCUS
control [4] to make it produce the finest and
distinctest definition.
5. Make sure tha t the traces align with horizontal graticule lines by
turning the CH I vertical
POSITION
control [ 17J. Where the traces do
not align with horiz
ont
al graticule lines, then make them align with
each other by adjusting the
TRA
CE ROTATION [SJ.
6,
Turn
the
HORIZONTAL
POSITION
control
[26]
to make it align with the
left-most graticule line.
7. Connect the PROB E to the
CHI
, X IN connector [9] to make it align
with the left-most graticule line.
8. If the top and a certain porti
on
of the square waves are tilted or
pointed, ad
ju
st the control terminal of the PROBE by means of a
small screwdriver as shown
on
Fig. 2-b(b ).
31