1-3. Points to be checked prior to use
Comply with the following procedures for safety and to prevent damage to
the product prior to operating this product.
1-3-1. Line voltage selection
Before use, check the voltage.
This instrument must be operated with the correct line voltage selector
switch setting to prevent damage in reference to table I-I.
To change the line voltage select
ion
:
1. Decide the voltage range and fuse in reference to table I-I.
2. Make sure the instrument is disc
onnect
ed from the power source.
3. Pull
out
the Line Voltage Selector switch on the rear
pan
el. Select
the arrow mark position of the switch from Table I-I. Slide the
arrow
mark
to the desired position
and
plug it in.
<Caution> : This product
has
the gro
un
d chassis (3 wire power cord is
used). Check whether any
other
equipment connecting with
this product requires the transformer bef
or
e use. If so, do
not connect the DC/AC or the hot chassis equipment if no
transformer is available.
Do not directly con nect it to the AC power
nor
to the
circuit directly
con
nected to the AC power.
Otherwise serious personal injury or
damag
e to this product
for a long time without trouble.
Table 1-1. Line voltage selection and fuse ratings
Line voltage Arrow mark position
Fuse Ratings(250V)
ULl98G
IECI27
90 - IIOVAC
100
2A F2A
108 - 132VAC
120
198 - 242 VAC
220
IA FIA
207 - 250 VAC 230